logo
Haus > produits >
Schaltkreislaufschalter-Analysator
>
Quantitative SF6 Leak Detector with Vacuum Pump

Quantitative SF6 Leak Detector with Vacuum Pump

Produktdetails:
Herkunftsort: Xi'an Shaanxi China
Markenname: XZH TEST
Zertifizierung: ISO CE
Modellnummer: XHJL306
Ausführliche Information
Herkunftsort:
Xi'an Shaanxi China
Markenname:
XZH TEST
Zertifizierung:
ISO CE
Modellnummer:
XHJL306
Anzeichen-Methode:
LCD-Digitalanzeige und akustische und visuelle Alarmsignale
Alarmbereich:
0,01 ppm bis 500 ppm
Hervorheben:

High Light

Hervorheben:

Quantitative SF6 leak detector

,

Small vacuum pump gas leak detector

,

Highly sensitive sulfur hexafluoride leak detector

Trading Information
Min Bestellmenge:
1 SATZ
Verpackung Informationen:
Holzverpackung
Lieferzeit:
5-8 Werktage
Zahlungsbedingungen:
T/T
Versorgungsmaterial-Fähigkeit:
3000 Einheiten/Jahr
Produkt-Beschreibung

Its working principle lies in the fact that varying concentrations of sulfur hexafluoride in the air lead to different ionization levels of the mixed gas exposed to high-frequency electromagnetic fields. The sulfur hexafluoride concentration is then determined by measuring the ionization degree.

Quantitative SF6 Leak Detector with Vacuum Pump 0

Performance Features
★ The instrument uses a small vacuum pump, and the probe and main unit are separate. The probe features characteristics such as being non-toxic and not producing toxic gases, and has an LCD display for gas concentration.
★ When the measured gas concentration is greater than or equal to the set value, it will automatically emit an alarm sound.

Quantitative SF6 Leak Detector with Vacuum Pump 1

Technical Parameters
☆ Measurement Range: 0.01ppm~500ppm  Response Time: Less than 3 seconds
☆ Indication Method: LCD digital display and audible and visual alarm signals
☆ Alarm Range: 0.01ppm~500ppm
☆ Vacuum Pump: Pumping speed 0.5 liters/second  Pipe length: 4 meters
☆ Operating Time: Continuous operation for no more than 4 hours
☆ Operating Conditions: AC 220V 50Hz  7℃-40℃  ≤85%